Gw-instek GDS-3000 Series GKT-100 deskew fixture User Manual Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Équipement Gw-instek GDS-3000 Series GKT-100 deskew fixture User Manual. GW Instek GDS-3000 Series GKT-100 deskew fixture User Manual User Manual [en] Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 1
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
Deskew Fixture
GKT-100
USER MANUAL
ISO-9001 CERTIFIED MANUFACTURER
GETTING STARTED
The GKT-100 deskew fixture is used to compensate for
the propagation delay between a passive voltage probe
and current probe. It is used with the GDS-3000.
Required tools
GDS-3000 x 1
GKT-100 x 1
USB type A-B cable x1 -used for deskew fixture
Standard passive probe x1
Current probe x1 (GCP-530 or GCP-1030)
CONNECT THE DESKEW FIXTURE
Before performing the manual deskew procedure,
please connect the deskew fixture to the USB host port
on the GDS-3000. Connect a passive voltage probe to
the JP1 terminal on the fixture and the current probe as
shown in the next picture.
1
3
2
4
Item
Descripon
1
Connect the USB cable: Connect the USB cable to the
USB port on the deskew fixture and the other end of
the cable to the USB host port on the GDS-3000.
2
Connect the current probe: Connect a current probe
(GCP-530 or GCP-1030) to the loop on the deskew
fixture and the other end of the current probe to the
CH1 terminal on the front panel of the GDS-3000.
3
Connect the passive probe: Connect the passive
probe to the JP2 terminal on the deskew fixture and
the other end to the CH2 terminal on the GDS-3000
front panel.
4
Grounding: Connect the grounding clip of the passive
voltage probe to JP1 terminal on the deskew fixture.
Note
Demagneze and zero the current probe. Refer
to the current probe’s documentaon for
instrucons on how to do this.
Make sure the direcon of the arrow on the
current probe and that on the current loop are
poinng in the same direcon.
PERFORMING DESKEW PROCEDURE
Follow the deskew procedure steps below once the
deskew fixture and probes are connected.
Step 1. Press the Autoset key on the
front panel of GDS-3000.
2. Press the CH1 key.
3. Press Probe from the bottom
menu.
4. Press Deskew on the side menu
and use the variable knob to set
the deskew time for the current
probe.
This manual contains proprietary information, which is
protected by copyright. All rights are reserved. No part
of this manual may be photocopied, reproduced or
translated to another language without prior written
consent of Good Will Corporation.
The information in this manual was correct at the time
of printing. However, Good Will continues to improve
its products and therefore reserves the right to change
the specifications, equipment, and maintenance
procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng
Dist., New Taipei City 236, Taiwan.
5. Press the CH2 key.
6. Repeat steps 3 to step 4 to adjust
the deskew time for the passive
voltage probe.
7. Adjust the waveform output of
voltage and current so that no
lead or lag is shown between
both waveforms to finish the
deskew adjustment.
SPECIFICATIONS
The specifications apply when the oscilloscope is
powered on for at least 30 minutes under +20°C~+30°C.
Model Specific Specifications
Temperature
Ambient temperature: -10 ~ 55
°
C
Storage Temperature: 20 ~60°C, no
condensaon
Relave humidity: 93% @40°C
65% @ 41°C~60°C
Storage: 90% @ 65 °C for 24 hr
Altitude
Operang: to 4,570 m (15,000)
Storage: to 15,244 m (50,000 )
Humidity
GW INSTEK PART NO. 82KT-10000M01
Vue de la page 0
1

Résumé du contenu

Page 1 - Deskew Fixture

Deskew FixtureGKT-100 USER MANUAL ISO-9001 CERTIFIED MANUFACTURER GETTING STARTED The GKT-100 deskew fixture is used to compensate for the propag

Commentaires sur ces manuels

Pas de commentaire