Gw-instek GDS-1000A-U Series Programming Manual Manuel d'utilisateur Page 83

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 88
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 82
COMMAND DETAILS
83
Example
:trigger:type 2
:trigger:pulse:time 1
Selects the pulse trigger
and sets the trigger
time as 1sec
:TRIGger:REJect
Set
Query
Description
Selects the trigger rejection filter.
Same as: Trigger key F4 F3
Syntax
< Long >
< Short >
:trigger:reject <NR1>
:trigger:reject?
:trig:rej <NR1>
:trig:rej?
Parameter
<NR1>
Rejection filter
0
off
1
LF
2
HF
Note
Before using this command, select the edge or
pulse trigger. See the example below.
Example
:trigger:type 0
:trigger:reject 1
Selects the edge trigger
and LF rejection filter
:TRIGger:SLOP
Set
Query
Description
Selects the trigger slope.
Same as: Trigger key F4 F1
Syntax
< Long >
< Short >
:trigger:slop <NR1>
:trigger:slop?
:trig:slop <NR1>
:trig:slop?
Parameter
<NR1>
Trigger slope
0
+ (positive)
1
(negative)
Vue de la page 82
1 2 ... 78 79 80 81 82 83 84 85 86 87 88

Commentaires sur ces manuels