Copyright 2003 GOODWILL INSTRUMENT CO., LTD.
GSP827 Command List
Note in examples
1. Returning message “OK” returns only for the software version V1.02 and below. (V0.94, V1.01).
2. Characters in italics represent the returning messages.
3. “^” means “space”.
4. Delay time is recommended only for reference. It could differ from different application.
5. All the Marker ( MEAS:MARK:xxxx) and Peak (MEAS:PEAK:xxxx) functions are available only for MM mode ON.
Command List <parameter> Delay
(ms)
Description
Parameters
(The unit is not case sensitive).
Example
(^ :space)
*CLS 300
Clear Status Command
*CLS
*ESE 300 Standard Event Status Enable
Command
*ESE
*ESE? 500
Standard Event Status Enable Query
*ESE?
*ESR? 500 Standard Event Status Register
Query
*ESR?
*IDN? 500 Identification Query *IDN?
*OPC 300 Operation Complete Command *OPC
*OPC? 500 Operation Complete Query *OPC?
*RST 300 Reset Command *RST
*SRE 300
Service Request Enable Command
*SRE
*SRE? 500 Service Request Enable Query *SRE?
*STB? 500 Read Status Byte Query *STB?
*TST? 500 Self-Test Query *TST?
*WAI 300 Wait-to-Continue Command *WAI
MEASure FREQuency CENter <value><unit
>
500 Set the center frequency Range: From 9k to 2.7GHz. Unit can be kHz,
MHz and GHz
MEAS:FREQ:CEN^100000^kHz
OK
MEASure FREQuency CENter? 500 Query the center frequency in kHz Reply: <value> <unit> MEAS:FREQ:CEN?
100000.0^kHz
MEASure FREQuency STart <value><unit
>
500 Set the start frequency Range: From 9k to 2.7GHz. Unit can be kHz,
MHz and GHz
MEAS:FREQ:ST^50^MHz
OK
MEASure FREQuency STart? 500 Query the start frequency in kHz MEAS:FREQ:STP?
100000.0^kHz