PSM SERIES PROGRAMMABLE POWER SUPPLY
PROGRAMMER MANUAL
57
Syntax:
CALibration:CURRent:MINimum <Decimal numbers>
CALibration:CURRent:MIDDle <Decimal numbers>
CALibration:CURRent:MAXimum <Decimal numbers>
CALibration:CURRent:PROTection
Function:
Auto calibrate OCP.
Syntax:
CALibration:CURRent:PROTection
CALibration:SAVe
Function:
To leave calibration mode after save.
Syntax:
CALibration:SAVe
CALibration:END
Function:
To leave calibration mode without saving.
Syntax:
CALibration:END
Note:
1) General calibration procedure:
Input Password to enter calibration mode, send command
CALibration:TYPe 0, then according to the value measured with
PSM SERIES PROGRAMMABLE POWER SUPPLY
PROGRAMMER MANUAL
58
voltage meter, use the commands of
CALibration:VOLTage:MINimum<measured value>,
CALibration:VOLTage:MIDDle < measured value >,
CALibration:VOLTage:MAXimum < measured value >
in sequence to complete the voltage calibration, then the program will
select type 1 automatically, therefore, just send the command
CALibration:VOLTage:PROTection directly, the OVP calibration
will be done
automatically.
The current and OCP calibration method is the same as above
procedure. After calibration, send the command CALibration:SAVe to
save the calibration value, then exit the calibration mode, or send the
command CALibration:end not to save the calibration value and exit
the calibration mode.
2) During calibrating, use the command CALibration:TYPe to select the
calibration type, the same type can be calibrated repeatedly.