GDS-1000 Series User Manual
72
Remote Control Interface
The Remote control interface section describes how to set up the
USB interface for PC connection. The details of remote control
commands are described in the GDS-1000 Programming Manual.
USB connection PC side
Type A, host
GDS-1000 side
Type B, slave
Speed
1.1/2.0 (full speed)
Procedure
1. Connect the USB cable to
the USB slave port.
2. When the PC asks for the USB driver, select
dso_cdc_1000.inf which is downloadable from
the GW website, www.gwinstek.com
, GDS-
1000 product corner.
3. On the PC, activate a terminal application such
as MTTTY (Multi-Threaded TTY). To check the
COM port No., see the Device Manager in the
PC. For WindowsXP, select Control panel →
System → Hardware tab.
4. Run this query command via the terminal
application.
*idn?
This command should return the manufacturer,
model number, serial number, and firmware
version in the following format.
GW, GDS-1022, 000000001, V1.00
5. Configuring the command interface is
completed. Refer to the programming manual
for the remote commands and other details.