GDS-1000 Series User Manual
94
7. The Channel1 calibration
will complete in less than 5
minutes.
Ch1 calibration 1/3
8. When finished, connect the
calibration signal to the
Channel 2 input and repeat
the procedure.
Done!!
9. The calibration is completed and the display
goes back to the previous state.
Probe Compensation
Procedure
1. Connect the probe between Channel1 input and
the probe compensation output (2Vp-p, 1kHz
square wave) on the front panel. Set the probe
attenuation to x10.
CH1
2. Press the Utility key.
Utility
3. Press ProbeComp.
4. Press Wavetype repeatedly to
select the standard square
wave.
Wave Type