Electrical Safety Tester
GPT-9000 Series
QUICK START GUIDE
GW INSTEK PART NO. 82PT-99040MA1
ISO-9001 CERTIFIED MANUFACTURER
Vue de la page 0
1 2 3 4 5 6 ... 19 20

Résumé du contenu

Page 1 - Electrical Safety Tester

Electrical Safety Tester GPT-9000 Series QUICK START GUIDE GW INSTEK PART NO. 82PT-99040MA1 ISO-9001 CERTIFIED MANUFACTURER

Page 2

GPT-9000 Series Quick Start Guide 10 Set the Test Voltage/Current (MANU mode only)  Use the UP/DOWN arrow keys to bring the cursor to the Test v

Page 3 - Table of Contents

OPERATION 11 Reference (MANU mode only)  Use the UP/DOWN arrow keys to bring the cursor to REF#.  Use the scroll wheel to set the reference. 

Page 4 - INTRODUCTION

GPT-9000 Series Quick Start Guide 12 Add a Test to the AUTO Test (AUTO mode only)  Use the DOWN arrow key to bring the cursor to MANU number. 

Page 5 - Panel Overview

OPERATION 13 WARNING If Double Action is active, the START button must be pressed 500ms after the STOP button was pressed to be able to start testi

Page 6 - OPERATION

GPT-9000 Series Quick Start Guide 14 Get Test Results  Wait for the MANU/AUTO test to finish. Abort MANU Test/AUTO Test  Press the STOP button

Page 7 - MANU number

OPERATION 15 AUTO MODE Results Overview: P A SP A SSSP A SF A ISLE MA U OT N A_ *A U T O = 0 0 1 - * *# 0 1 : # 0 2 : # 0 3 : # 0 4 : P A S

Page 8 - MANU mode

GPT-9000 Series Quick Start Guide 16 AUTO MODE P A SP A SSSP A SF A ISLE MA U OT N A_ *A U T O = 0 0 1 - * *# 0 1 : # 0 2 : # 0 3 : # 0

Page 9

OPERATION 17 Move Steps  Use the arrow keys to highlight the step you wish to move.  Press the MOVE soft-key.  Use the arrow keys to move the

Page 10 - I M ET R

GPT-9000 Series Quick Start Guide 18 The step will be removed from the table as soon as the DEL soft-key is pressed. Save Page View  To save, p

Page 11 - EDIT/SAVE

OPERATION 19 Select a Utility Setting.  Choose a utility by pressing the corresponding LCD, BUZZ, INTER or CTRL soft-key.  The chosen utility wi

Page 12 - READY Status

This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied,

Page 13 - TEST Status

GPT-9000 Series Quick Start Guide 20 MANU Utility Settings Description The MANU Utility menu is accessed by pressing the UTILITY key when the tes

Page 14 - Results

Table of Contents 3 Table of Contents INTRODUCTION ... 4 GPT-9000 Series Overview ...

Page 15 - STOP Status

GPT-9000 Series Quick Start Guide 4 INTRODUCTION This Quick Start Guide is intended as a fast introduction to operating the GPT-9000 Series Safety

Page 16 - Page View

INTRODUCTION 5 Panel Overview Front Panel (GPT-9801/9802/9803) POWERSTART STOPREMOTEMANU/AUTO EDIT/SAVE UTILITYESC PAGEGPT-9803TesterHIGH VOLTAGECA

Page 17 - D E L

GPT-9000 Series Quick Start Guide 6 OPERATION Menu Tree This section describes the overall structure of the operation statuses and modes for the

Page 18 - Common Utility Settings

OPERATION 7 VIEW Status Description VIEW status is used to select a manual test in the MANU mode or an automatic test/step in AUTO mode. VIEW statu

Page 19 - L C D

GPT-9000 Series Quick Start Guide 8 Go to EDIT Status  Press the EDIT/SAVE key. EDIT/SAVE Go to READY Status  Press the STOP button when a MA

Page 20 - MANU Utility Settings

OPERATION 9 EDIT Status Description EDIT status is accessed by pressing the EDIT/SAVE key when in VIEW status. EDIT status is used to edit the curr