Gw-instek GSP-827 Programming Manual Manuel d'utilisateur Page 10

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 18
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 9
Copyright 2003 GOODWILL INSTRUMENT CO., LTD.
CONtrol RBW MANual 200 Assign the RBW in manual mode CON:RBW:MAN
OK
CONtrol RBW MODE? 200 Query the RBW mode. Auto, Manual CON:RBW:MODE?
Manual
CONtrol RBW? 200 Query the RBW CON:RBW?
30 kHz
CONtrol VBW <index> 400 Assign the VBW by index number 0=10Hz, 1=30Hz, 2=100Hz, 3=300Hz, 4=1kHz,
5=3kHz, 6=10kHz, 7=30kHz, 8=100kHz,
9=300kHz, 10=1MHz
CON:VBW^4
OK
CONtrol VBW AUTO 400 Assign the VBW in auto mode CON:VBW:AUTO
OK
CONtrol VBW MANual 200 Assign the VBW in manual mode CON:VBW:MAN
OK
CONtrol VBW MODE? 700 Query the VBW mode. CON:VBW:MODE?
Manual
CONtrol VBW? 600 Query the VBW CON:VBW?
1 Khz
CONtrol SYStem GPIB ADDRess
<value>
300 Assign the GPIB address 1 to 30 CON:SYS:GPIB:ADDR^9
OK
CONtrol SYStem GPIB ADDRess? 500 Query the GPIB address CON:SYS:GPIB:ADDR?
9
CONtrol SYStem CONTrast <value> 200 Assign the contrast number of the
LCD module
128(white)~176(black) CON:SYS:CONT^145
OK
CONtrol SYStem CONTrast? 400 Query the contrast number of LCD
module.
CON:SYS:CONT?
145
CONtrol SYStem CLOCK? 100 Query the system date and clock CONtrol:SYStem:CLOCK?
23:53:00
Monday May 26, 2003
CONtrol SYStem FirmWare? 700 Query the Firmware version CONtrol:SYStem:FirmWare?
Firmware version = 2.0
CONtrol SYStem VERsion? 600 Query the Software version and built
date
CONtrol:SYStem:VERsion?
System Software Version = 00.94
03/31/03
CONtrol SYStem PRESET 200 Preset the system CON:SYS:PRESET
OK
CONtrol SYStem DISPlay <value> 100 Set the display mode. 0:Normal,
1:Inverse
CON:SYS:DISP^0
OK