FAQ
97
The probe waveform is distorted.
You might need to compensate the probe. For details, see page94.
Note that the frequency accuracy and duty factor are not specified
for probe compensation waveform and therefore it should not be
used for other reference purpose.
Autoset does not catch the signal well.
Autoset function cannot catch signals under 30mV or 30Hz. Please
use the manual operation. See page37 for details.
I want to clean up the cluttered panel settings.
Recall the default settings by pressing the Save/Recall keyDefault
Setting. For default setting contents, see page34.
The saved display image is too dark on the background.
Use the Inksaver function which reverses the background color. For
details, see page84.
The accuracy does not match the specifications.
Make sure the device is powered on for at least 30 minutes, within
+20°C~+30°C. This is necessary to stabilize the unit to match the
specification.
The SD card slot does not accept my card.
Make sure it is a 1. Standard SD card (MMC and SDHC are not
supported), 2. 2GB or less, and 3. FAT or FAT32 formatted.
It is recommended to use other SD card readers stead, e.g. San
Disk, Kingston, Transcend, once the SD Card readers has an
Vue de la page 96