Differential Probe GDP-025/050/100 Series USER MANUAL GW INSTEK PART NO. 82DP-10000MA1 ISO-9001 CERTIFIED MANUFACTURER
GDP Series User Manual 10 black x1) Accessories GDP-050 Adapter (110V or 220V depending on region) BNC to BNC cable: 50Ω resistance, RG58C U
GETTING STARTED 11 Instrument Overview GDP-025 Intput terminalsOver range indicatorAttenuation function switchPower indicatorPower supply inputOutp
GDP Series User Manual 12 Input terminals Category III input terminals. Power Indicator The power indicator lights when the external power suppl
MEASUREMENT 13 MEASUREMENT High voltage Risk of electric shock. The GDP Series differential voltage probes are designed for use with high voltages
GDP Series User Manual 14 Caution The power light must come on Ensure the range light does not come on before measuring Deviation Table Mod
MEASUREMENT 15 Example With the probe set to a factor of X200, the oscilloscope set to a vertical deviation of 0.5V/div, the real vertical deviati
GDP Series User Manual 16 APENDIX GDP Series Specifications GDP-025 Bandwidth DC-25MHz (-3dB)for x50, x200 DC-50MHz (for attenuation x20) At
APENDIX 17 Electrical Safety to IEC 1010-1 Dual Insulation Installation Category III Degree of Pollution 2 Rated Voltage or Max Line-Earth: 600Vrm
GDP Series User Manual 18 GDP-050 Bandwidth DC-50MHz (-3dB)for x200, x500, x1000 DC-25MHz (for attenuation x100) Attenuation X100, x200, x50
APENDIX 19 GDP-100 Bandwidth DC-100MHz (-3dB)for x200, x500, x1000 DC-50MHz (for attenuation x100) Attenuation X100, x200, x500, x1000 Accur
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied
GDP Series User Manual 20 EC Declaration of Conformity We GOOD WILL INSTRUMENT CO., LTD. No.7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 23
APENDIX 21
TABLE OF CONTENTS 3 Table of Contents SAFETY INSTRUCTIONS ... 4Safety Symbols ...
GDP Series User Manual 4 SAFETY INSTRUCTIONS This chapter contains important safety instructions that should be followed when operating and storin
SAFETY INSTRUCTIONS 5 Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the sup
GDP Series User Manual 6 Electrical Safety WARNING Make sure the maximum differential voltage does not exceed 1400V (DC+AC peak) or 450 Vrms f
SAFETY INSTRUCTIONS 7 Storage environment Location: Indoor Relative Humidity: 10 to 90% Temperature: -30°C to 70°C Note: If the probe is n
GDP Series User Manual 8 Power cord for the United Kingdom When using the instrument in the United Kingdom, make sure the power cord meets the fol
GETTING STARTED 9 GETTING STARTED The Getting Started chapter introduces the probe’s features, appearance and functions. Main Features Model name