COMMAND DETAILS
13
*RST
Set
Description
Resets the GDS-1000-U (recalls the default panel
settings).
Same as: Save/Recall key → F1
Syntax
*rst
:SYSTem:ERRor
Query
Description
Returns the oscilloscope system error messages, if
there are any.
Syntax
< Long >
< Short >
:system:error?
:syst:err?
Parameter
ID
Contents
ID
Contents
-100
command error
-102
syntax error
-220
parameter error
-221
settings conflict
-222
data out of range
-223
too much data
-224
illegal parameter
-232
invalid format
Example
:system:error?
-102
Indicates that the
command syntax is
wrong
:SYSTem:VERSion
Query
Description
Returns the oscilloscope firmware version.
Same as: Utility key F4 (only the firmware
version)
Syntax
< Long >
< Short >
:system:version?
:syst:vers?
Note
For retrieving all system information including the
firmware version, use the *idn? command.
Vue de la page 12