Gw-instek GDS-1000-U Series Programming Manual Manuel d'utilisateur Page 51

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 70
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 50
COMMAND DETAILS
51
:DATALOG:SAVe
Set
Query
Description
Sets the save type as waveform or image.
Syntax
< Long >
< Short >
:DATALOG:SAVe {0|1}
:DATALOG:SAVe?
:DATALOG:SAV {0|1}
:DATALOG:SAV?
Parameter/
Return parameter
0
Save as image
1
Save as waveform
Example
:DATALOG:SAVe 1
Set the save type to
waveform.
:DATALOG:INTerval
Set
Query
Description
Sets or queries the interval time between each
recording.
Syntax
< Long >
< Short >
:DATALOG:INTerval <NR1>
:DATALOG:INTerval?
:DATALOG:INT
<NR1>
:DATALOG:INT?
Parameter/
Return parameter
<NR1>
Discrete time intervals in seconds:
{2|3|4|5|10|20|30|60|120|300|600|1
200|1800}
Example
:DATALOG:INT 2
Sets the interval time
to 2 seconds.
:DATALOG:DURation
Set
Query
Description
Sets or queries the duration time of each recording.
Syntax
< Long >
< Short >
Vue de la page 50
1 2 ... 46 47 48 49 50 51 52 53 54 55 56 ... 69 70

Commentaires sur ces manuels