COMMAND DETAILS
43
Note
Before any Go-NoGo command can be used,
please use the :GONogo:FUNCtion 1 command to
initialize the oscilloscope.
Syntax
< Long >
< Short >
:GONogo:SOURce {1|2}
:GONogo:SOURce?
:GON:SOUR {1|2}
:GON:SOUR?
Parameter/
Return parameter
1
Sets the source to channel 1
2
Sets the source to channel 2
Example
:GON:SOUR 1
Sets the source to
channel 1.
:GONogo:VIOLation
Set
Query
Description
Sets or queries the Go-NoGo violation conditions.
Note
Before any Go-NoGo command can be used,
please use the :GONogo:FUNCtion 1 command to
initialize the oscilloscope.
Syntax
< Long >
< Short >
:GONogo:VIOLation {0|1}
:GONogo:VIOLation?
:GON:VIOL {0|1}
:GON:VIOL?
Parameter/
Return parameter
0
Violation condition = “Continue”
1
Violation condition = “Stop”
Example
:GON:VIOL 1
Sets the violation
condition to
“Continue”.
:TEMPlate:MODe
Set
Query
Description
Sets or queries the Go-NoGo template mode.
When Auto mode is selected, CH1 or CH2 are
used as the template source. When Normal mode
is selected, the template source can be selected
from internal memory (W1~W15, RefA or RefB).