:GONogo:EXECute
Start or stop the execution of GO/NO-GO comparison function. This command is
equal to the function key “F4” of GO/NO-GO menu.
Syntax
:GONogo:EXECute {0|1}
:GONogo:EXECute?
Arguments
0Stop GO/NO-GO comparison 1Start GO/NO-GO comparison
Returns
<NR1>
Note: Before running any Go/NoGo commands, the oscilloscope has to be initialized, please
add the command of “:GONogo:FUNCtion 1 in order to initial the oscilloscope.
:GONogo:FUNCtion
This command is used to initialize or relieve the GO/NO-GO function. Before the
GO/NO-GO comparison is started. This command must be executed to enable the
comparison unit.
Syntax
:GONogo:FUNCtion {0|1}
:GONogo:FUNCtion?
Arguments
0Relieve GO/NO-GO function
1Initialize GO/NO-GO function
Returns
<NR1>