GDS-2000 Series Programming Manual
70
:TEMPlate:MIN
Sets and query the number of template used for minimum boundary.
Before running this command, the oscilloscope has to be initialized as “Normal
Template” mode, please add the command of “:TEMPlate:MODe 0 in order to
change the settings.
Syntax
:TEMPlate:MIN {0|20}
:TEMPlate:MIN?
Arguments
0 Ref B
1~20 Internal memory W1~W20
Returns
<NR1>
Note: A template can be defined as either MAX or MIN template only.
Note B: Before running this command, the oscilloscope has to be initialized, please add the
command of “:GONogo:FUNCtion 1 in order to initial the oscilloscope.