Gw-instek LCR-800 Series Manuel d'utilisateur Page 60

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 72
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 59
INTERACE
119
Handler Timing
Background
The handler timing characteristics are described in
the timing diagram. Times T1 to T6 are described
in the relevant tables.
Timing Characteristics
Trigger Pulse Width T1 MIN MAX
5us ~
Measurement start delay
time
T2 MIN MAX
140us ~
/O_EOM Delay Time After
Data Output
T3 MIN MAX
5us ~
Calculation and binning time T4 MIN MAX
6ms ~
LCR-800 User Manual
120
Analog
Measurement
time
T5 Slow Medium Fast
0.012kHz 817ms 817ms 817ms
0.1kHz 901ms 125ms 125ms
0.12kHz 901ms 105ms 103ms
1kHz 903ms 59ms 27ms
10kHz 873ms 53ms 17ms
100kHz 873ms 53ms 17ms
Trigger Wait Time
After /O_EOM
Output
T6 Slow Medium Fast
OFF 2ms 2ms 2ms
BIN 4ms 4ms 4ms
VALUE 16ms 16ms 16ms
Binning Accuracy
Fast Medium Slow
LCR_827 0.5% 0.2% 0.1%
LCR_829 0.5% 0.2% 0.1%
LCR_826 0.5% 0.2% 0.1%