Gw-instek PSU-Series Programming User Manual Manuel d'utilisateur Page 71

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 128
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 70
REMOTE CONTROL
71
:OUTPut[:STATe]:TRIGgered
Set
Query
Description
Turns the output on or off when a software trigger
is generated.
Syntax
Query Syntax
:OUTPut[:STATe]:TRIGgered { <bool>|OFF|ON }
:OUTPut[:STATe]:TRIGgered?
Parameter
OFF | 0
Turns the output off when a software trigger
is generated (*TRG).
ON | 1
Turns the output on when a software trigger
is generated (*TRG).
Return parameter
<bool>
Returns output trigger status of the
instrument.
:OUTPut:PROTection:CLEar
Set
Description
Clears over-voltage, over-current and over-
temperature (OVP, OCP, OTP) protection circuits.
It also clears the shutdown and sense protection
circuit. The AC failure protection cannot be
cleared.
Syntax
:OUTPut:PROTection:CLEar
Vue de la page 70