AFG-2225 Series User Manual
242
SOURce[1|2]:FREQuency:STARt
Source Specific
Command
Sets the start frequency of the sweep. 100Hz is the
default start frequency.
To sweep up or down, set the stop frequency
higher or lower than the start frequency.
SOURce[1|2]:FREQuency:STARt
{<frequency>|MINimum|MAXimum}
<frequency>
1μHz~ 25MHz
1μHz~ 15MHz(Square)
1μHz~ 1MHz (Ramp)
SOUR1:FREQ:STAR +2.0000E+03
Sets the start frequency to 2kHz.
SOURce[1|2]:FREQuency:STARt? [MINimum|
MAXimum]
<NR3>
Returns the start frequency
in Hz.
SOUR1:FREQ:STAR? MAX
+8.0000E+0
Returns the maximum start frequency allowed.
SOURce[1|2]:FREQuency:STOP