Gw-instek GDS-200 Series Programming Manual Manuel d'utilisateur Page 54

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 139
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 53
GDS-200 & GDS-300 Programming Manual
54
:DISPlay:PERSistence
Description
Sets or queries the waveform persistence level.
Syntax
:DISPlay:PERSistence { INFInite | SHORT | MEDium |
LONG | OFF| ? }
Parameter
INFInite
Infinite persistence
SHORT
Shortest persistence time
MEDium
Medium persistence
LONG
Long persistence
OFF
No persistence
Return Parameter
Returns the persistence setting.
Example
:DISPlay:PERSistence LONG
Sets the persistence the longest time.
:DISPlay:GRATicule
Description
Sets or queries graticule display type.
Syntax
:DISPlay:GRATicule {FULL | GRID|CROSs | FRAMe | ?}
Parameter
FULL
CROSs
FRAMe
GRID
Return parameter
Returns the graticule type.
Example
:DISPlay:GRATicule FULL
Sets the graticule to .
:DISPlay:WAVEform
Description
Sets or queries whether the waveforms are drawn
as vectors or dots.
Set
Query
Set
Query
Set
Query