Gw-instek GDS-200 Series Programming Manual Manuel d'utilisateur Page 63

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 139
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 62
COMMAND DETAILS
63
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the global low value.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:LOW?
1.00E-0
Selects Channel 1, and then measures the low
current/voltage.
:MEASure:MEAN
Description
Returns the mean voltage/current of one or more
full periods.
Syntax
:MEASure:MEAN?
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the mean.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:MEAN?
1.82E-3
Selects Channel 1, and then measures the mean
value.
Query
Vue de la page 62