COMMNAND OVERVIEW
37
Input State Commands
:INPut ..................................................................................... 37
:INPut:SHORt ...................................................................... 37
:INPut:MODE ...................................................................... 38
:INPut
Set
Query
Description
Sets or queries the status of the load.
Syntax
:INPut {<Boolean>|OFF | ON }
Query Syntax
:INPut?
Parameter
OFF or 0
Load = OFF
ON or 1
Load = ON
Return parameter
0
Load = OFF
1
Load = ON
Example
:INPut ON
Turns the load on.
Query example
:INPut?
>0
Indicates that the load is off.
:INPut:SHORt
Set
Query
Description
Shorts or opens the input terminals or queries their
status.
Syntax
:INPut:SHORt {<Boolean>|OFF | ON }
Query Syntax
:INPutSHORt?
Parameter
OFF or 0
Short = OFF
ON or 1
Short = ON
Return parameter
0
Short = OFF
1
Short = ON
Example
:INPut:SHORt ON
Shorts the input terminals.