COMMNAND OVERVIEW
71
Query example
:CURRent:VA?
>1.0A
Returns the “A Value” current setting.
:CURRent:VB
Set
Query
Description
Sets or queries the CC mode “B Value” (static
mode) current. This command is applicable to
static modes.
Note: A different current value can be set for each
current range (High/Mid/Low).
Syntax
:CURRent:VB {<NRf>[ A ] | MINimum | MAXimum }
Query Syntax
:CURRent:VB?
Parameter
<NRf>[A]
“B Value” current value
MINIMUM or MIN
Minimum current level
MAXIMUM or MAX
Maximum current level
Return parameter
<NR2>
Returns the “B Value”
current value.
Example
:CURRent:VB MIN
Sets the current value to the minimum.
Query example
:CURRent:VB?
>1.0A
Returns the “B Value” current setting.
:CURRent:SRATe
Set
Query
Description
Sets or queries the current slew rate for CC static
mode.
Syntax
:CURRent:SRATe {<NRf>| MINimum | MAXimum }
Query Syntax
:CURRent:SRATe?
Parameter
<NRf>
Sets the slew rate in
mA/us
MINIMUM or MIN
Set to the highest slew rate
MAXIMUM or MAX
Set to the lowest slew rate