COMMNAND OVERVIEW
81
Syntax
:RESistance:SRATe {<NRf> | MINimum | MAXimum}
Query Syntax
:RESistance:SRATe?
Parameter
<NRf>
Sets the slew rate in
mA/us
MINIMUM or MIN
Set to the lowest slew rate
MAXIMUM or MAX
Set to the highest slew rate
Return parameter
<NR2>
Returns the slew rate in
mA/us.
Example
:RESistance:SRATe MIN
Sets the slew rate to the minimum.
Query example
:RESistance:SRATe?
>5.0000
Returns the slew rate as 5mA/us.
:RESistance:L1
Set
Query
Description
Sets or queries the CR mode “Level1”
conductance. This command is only applicable to
dynamic mode.
Note: A different value can be set for each current
range (High/Mid/Low).
Syntax
:RESistance:L1 {<NRf>[ MHO ] | MINimum |
MAXimum }
Query Syntax
:RESistance:L1?
Parameter
<NRf>[ MHO ]
level1” conductance
value (CR Unit: MHO )
MINIMUM or MIN
Minimum conductance
level
MAXIMUM or MAX
Maximum conductance
level
Return parameter
<NR2>
Returns the “level1”
conductance value.
Example
:RESistance:L1 MIN
Sets the conductance value to the minimum.