68
Parameter
HIGH
High range
MIDDLE or MIDD
Middle range
LOW
Low range
Return parameter
High
High range
Mid
Middle range
Low
Low range
Example
:CRANge LOW
Sets the current range to Low.
Query example
:CRANge?
>Low
The current range is set to Low.
[:MODE]:VRANge
Set
Query
Description
Configures the voltage range for all the applicable
operating modes.
Syntax
[:MODE]:VRANge {HIGH | LOW }
Query Syntax
[:MODE]:VRANge?
Parameter
HIGH
High range
LOW
Low range
Return parameter
High
High range
Low
Low range
Example
:VRANge LOW
Sets the voltage range to Low.
Query example
:VRANge?
>LOW
The voltage range is set to Low.
[:MODE]:RESPonse
Set
Query
Description
Returns the CV Mode response speed. Fast is the
default response speed.
Syntax
[:MODE]:RESPonse { FAST | SLOW }
Query Syntax
[:MODE]:RESPonse?