REMOTE INTERFACE
205
SOURce[1|2]:APPLy:USER [<frequency> [,<amplitude>
[,<offset>] ]]
<frequency>
1μHz~60MHz
<amplitude>
1mVpp~10Vpp (50)
<offset>
-4.99V~4.99V (50)
SOUR1:APPL:USER
SOURce[1|2]:APPLy?
Source Specific
Command
Outputs a string with the current settings.
The string can be passed back appended to the
Apply Command.
SOURce[1|2]:APPLy?
<string>
Function, frequency,
amplitude, offset
SOUR1:APPL?
SIN +5.0000000000000E+03,+3.0000E+00,-2.50E+00
Returns a string with the current function and
parameters, Sine, 5kHz, 3 Vpp, -2.5V offset.