Gw-instek AFG-2225 User Manual Manuel d'utilisateur Page 283

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 301
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 282
REMOTE INTERFACE
283
-222 Data out of range;marker confined to sweep span; value clipped
to ...
Example: The marker frequency is set to a value
outside the start or stop frequencies. The marker
frequency is automatically adjusted to either the
start or stop frequency (whichever is closer to the set
value).
-222 Data out of range;FM deviation; value clipped to ...
Example: The frequency deviation is outside of
range. The deviation is automatically adjusted to an
upper or lower limit, depending on the frequency.
-222 Data out of range;trigger delay; value clipped to upper limit
Example: The trigger delay was set to a value out of
range. The trigger delay has been adjusted to the
maximum (655350 nseconds).
-222 Data out of range; trigger delay limited by length of burst; value
clipped to upper limit
Example: The trigger delay and the burst cycle time
combined must be less than the burst period.
-222 Data out of range;duty cycle; value clipped to ...
Example: The duty cycle is limited depending on the
frequency.
Duty Cycle
Frequency
50%
> 25MHz
10%~90%
100 KHz ~ 1MHz
1%~99%
< 100KHz
-222 Data out of range; duty cycle limited by frequency; value clipped to
upper limit
Example: The duty cycle is limited depending on the
frequency. When the frequency is greater than 50
MHz, the duty cycle is automatically limited to 50%.