Vue de la page 104
COMMAND DETAILS
105
Syntax
:MEASUrement:MEAS<X>:STATE { ON | OFF | 1 | 0
| ? }
Related
commands
:MEASUrement:MEAS<X>:SOUrce<X>
:MEASUrement:MEAS<X>:TYPe
Parameter
MEAS<X>
The automatic measurement number
from 1 to 8.
ON/1
Turn the measurement on.
OFF/0
Turn the measurement off.
Return parameter
0
Measurement is off.
1
Measurement is on.
Example
:MEASUrement:MEAS1:STATE 1
Turns measurement 1 on.
:MEASUrement:MEAS<X>:VALue
Description
Returns the measurement results for the selected
measurement. This is a statistics related command.
Syntax
:MEASUrement:MEAS<X>:VALue?
Related
Commands
:MEASure:SOURce<X>
Return parameter
MEAS<X>
The automatic measurement number
from 1 to 8.
Note
The measurement source(s), measurement
number, measurement type and measurement
state must first be set before a measurement result
can be returned.
Query