Gw-instek GDS-2000A series Programming Manual Manuel d'utilisateur Page 106

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 272
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 105
GDS-2000A Programming Manual
106
Example
:MEASUrement:MEAS1:SOUrce1 CH1
:MEASUrement:MEAS1:TYPe PK2PK
:MEASUrement:MEAS1:STATE ON
:MEASUrement:MEAS1:VALue?
5.000E+0
Selects channel 1 as the source for measurement 1,
sets measurement 1 to peak to peak measurement
and then turns on the measurement. The result
returns the peak to peak measurement.
:MEASUrement:MEAS<X>:MAXimum
Description
Returns the maximum measurement results for the
selected measurement from the last time the
statistics were reset. This is a statistics related
command.
Syntax
:MEASUrement:MEAS<X>:MAXimum?
Related
Commands
:MEASUrement:STATIstics:MODe
Parameter
MEAS<X>
The automatic measurement number
from 1 to 8.
Example
:MEASUrement:MEAS3:SOUrce1 CH1
:MEASUrement:MEAS3:TYPe PK2PK
:MEASUrement:MEAS3:STATE ON
:MEASUrement:STATIstics:MODe ON
:MEASUrement:MEAS3:MAXimum?
2.800E-02
Returns the maximum measurement result for
measurement number 3.
Query
Vue de la page 105
1 2 ... 101 102 103 104 105 106 107 108 109 110 111 ... 271 272

Commentaires sur ces manuels

Pas de commentaire