Gw-instek GDS-3000 Series Programming Manual Manuel d'utilisateur Page 74

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 103
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 73
GDS-3000 Programming Manual
74
:REF<x>:OFFSet
Set
Query
Description
Sets or returns the selected reference waveform
vertical position (offset).
Syntax
:REF<x>:OFFSet { <NRf> | ?}
Parameter
<x>
Reference
waveform
<NRf>
1,2,3,4
REF1~REF4
Vertical offset
Return parameter
Returns the reference waveform vertical position
as <NR3>.
Example
:REF1:OFFSet -5.000E-2
Selects reference 1, and then sets the vertical
position to -50mV/mA.
:REF<x>:SCALe
Set
Query
Description
Sets or returns the selected reference waveform
vertical scale.
Syntax
:REF<x>:SCALe { <NRf> | ?}
Parameter
<x>
Reference
waveform
<NRf>
1,2,3,4
REF1~REF4
Vertical scale
Return parameter
Returns the reference waveform vertical scale as
<NR3>.
Example
:REF1:SCALe 5.000E-2
Selects reference 1, and then sets the vertical scale
to 50mV|mA/div.
Vue de la page 73
1 2 ... 69 70 71 72 73 74 75 76 77 78 79 ... 102 103

Commentaires sur ces manuels

Pas de commentaire