GDS-3000 Programming Manual
90
:TRIGger:ALTernate
Set
Query
Description
Sets alternating between source triggers on or off.
Syntax
:TRIGger:ALTernate {OFF | ON |?}
Parameter
OFF
Alternate off
ON
Alternate on
Return parameter
Returns the Alternate trigger status (ON, OFF).
Example
:TRIGger:ALTernate ON
Turns on alternating between source triggers.
:TRIGger:LEVel
Set
Query
Description
Sets or queries the trigger level.
Syntax
:TRIGger:LEVel {<NRf> | ?}
Parameter
<NRf>
Trigger level value
Return parameter
Returns the trigger level as <NR3>.
Example
:TRIGger:LEVel 3.30E-3
Sets the trigger level to 330mV/mA.
Note: This command is equal to :TRIGger:HLEVel
:TRIGger:TIMEOut:WHEn
Set
Query
Description
Sets or queries the timeout trigger triggering point.
Note: Only applicable for 500MHz models with the
Timeout trigger APP installed.
Syntax
:TRIGger:TIMEOut:WHEn {HIGH | LOW | EITher | ?}
Related
commands
:TRIGger:TIMEOut:TIMER
Parameter/
Return parameter
HIGH
Sets the triggering point as stays high.
LOW
Sets the triggering point as stays low.
Vue de la page 89