Gw-instek GSP-930 Programming Manual Manuel d'utilisateur Page 102

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 195
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 101
GSP-930 Programming Manual
100
:DISPlay:DEMod[:WINDow]:TRACe:Y
[:SCALe]:RPOSition
Set
Query
Sets or queries the Reference Position of the trace
for AM/FM demodulation (y-axis grid division).
:DISPlay:DEMod[:WINDow]:TRACe:Y[:SCALe]:
RPOSition <integer>
:DISPlay:DEMod[:WINDow]:TRACe:Y[:SCALe]:
RPOSition?
<integer>
<NR1>1~10
:DISP:DEM:TRAC:Y:RPOS 2
:DISPlay:DEMod[:WINDow]:TRACe:Y
[:SCALe]:RVALue
Set
Query
Sets or queries the Reference value
(AM: %, FM: Hz).
:DISPlay:DEMod[:WINDow]:TRACe:Y[:SCALe]:
RVALue <NRf>
:DISPlay:DEMod[:WINDow]:TRACe:Y[:SCALe]:
RVALue?
<NRf>
AM Unit: %, FM Unit: Hz
<NR3>
AM Unit: %, FM Unit: Hz
:DISP:DEM:TRAC:Y:RVAL 2 %
:DISPlay[:WINDow]:NORMal
Set
Sets the display window to the normal trace mode.
:DISPlay[:WINDow]:NORMal
:DISP:NORM
Vue de la page 101