REMOTE CONTROL
67
Query Syntax
:CALCulate:ACPR:CHPower?
Return parameter
<NR3>
Power
Query Example
:CALC:ACPR:CHP?
>-1.028e+02
:CALCulate:ACPR:STATe
Set
Query
Description
Sets or queries the state of the ACPR measurement
function.
Syntax
:CALCulate:ACPR:STATe {OFF|ON|0|1}
Query Syntax
:CALCulate:ACPR:STATe?
Parameter
0
ACPR is disabled.
1
ACPR is enabled.
OFF
ACPR is disabled.
ON
ACPR is enabled.
Return parameter
0
ACPR is disabled.
1
ACPR is enabled.
Query Example
:CALC:ACPR:STAT?
>1
:CALCulate:CNR:RESult?
Query
Description
Returns the CNR measurement result in dB.
Query Syntax
:CALCulate:CNR:RESult?
Return parameter
<NR3>
CNR measurement in dB
Query Example
:CALC:CNR:RES?
>-4.959e+01
:CALCulate:CNR:STATe
Set
Query
Description
Sets or queries the state of the CNR measurement
function.
Syntax
:CALCulate:CNR:STATe {ON|OFF|1|0}
Vue de la page 68