Gw-instek GSP-930 Programming Manual Manuel d'utilisateur Page 67

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 195
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 66
REMOTE CONTROL
65
:CALCulate:ACPR:ACHannel<n>:LOWer?
Query
Description
Returns the ACPR (adjacent channel power ratio in
dB) calculated for the selected lower adjacent
channel.
Query Syntax
:CALCulate:ACPR:ACHannel<n>:LOWer?
Parameter
<n>
<NR1>adjacent channel 1~3
Return parameter
<NR3>
Power ratio in dB
Query Example
:CALC:ACPR:ACH1:LOW?
>1.801e+01
:CALCulate:ACPR:ACHannel<n>:UPPer?
Query
Description
Returns the ACPR (adjacent channel power ratio in
dB) calculated for the selected higher adjacent
channel.
Query Syntax
:CALCulate:ACPR:ACHannel<n>:UPPer?
Parameter
<n>
<NR1>adjacent channel 1~3
Return parameter
<NR3>
Power ratio in dB
Query Example
:CALC:ACPR:ACH1:UPP?
>1.921e+01
:CALCulate:ACPR:ACHannel<n>:STATe
Set
Query
Description
Sets or queries the state of the selected adjacent
channel.
Syntax
:CALCulate:ACPR:ACHannel<n>:STATe {OFF|ON|0|1}
Query Syntax
:CALCulate:ACPR:ACHannel<n>:STATe?
Parameter
<n>
<NR1>adjacent channel 1~3
0
Disable the selected channel.
1
Enable the selected channel.
OFF
Disable the selected channel.
Vue de la page 66
1 2 ... 62 63 64 65 66 67 68 69 70 71 72 ... 194 195

Commentaires sur ces manuels

Pas de commentaire