6.3 Command Descriptions
APS-1102 Programmable AC/DC Power Source 6-40
RST
This resets the device to its factory settings. However, reset cannot be executed when output is on.
([1, “Invalid with output on”] error occurs.)
Syntax: RST
Parameters: None
Command example: RST
Response format: None
Response example: None
SAV
This stores the current status information in memory. The memory storage ranges from 1 to 30.
Syntax: SAV <NR1>
Parameters: 1 to 30
Command example: SAV 1
Response format: None
Response example: None
SRE
This sets the service request enable register. The setting ranges from 0 to 178.
Syntax: SRE <NR1>
Parameters: 0 to 178
Command example: SRE 178
Query example: SRE?
Response format: <NR1>
Response example: 178
STB?
This returns the status byte register value. The response ranges from 0 to 178.
Syntax: STB?
Parameters: None
Query example: STB?
Response format: <NR1>
Response example: 2