BASIC MEASUREMENT
39
PPM will be displayed on the right hand side of
the screen, next to mode.
Set constant voltage source
Background
If a DUT needs to be tested at a set voltage, the
constant voltage function can be used. Using the
C.V. function the LCR will maintain a source
resistance of 25. Therefore the test voltage is
constant for any DUT impedance greater than 25.
Using the constant voltage feature will reduce the
accuracy of measurements by a factor of 3.
Panel operation
1. Press 1/C.V to turn constant
voltage on or off.
C.V ON / OFF is toggled each time the 1/C.V
button is pressed.
LCR-800 User Manual
40
Set Range hold
Background
When DUTs are disconnected from the test
cables/fixtures during continuous testing, Range
Hold can be used to avoid range switching. This is
particularly useful for repetitively testing a
number of DUTs. For more information on Range
and range hold, see the specifications, page 136.
Panel operation
1. Press 0/R.H to turn Range Hold on
or off.
R.H ON / OFF is toggled each time the 0/R.H
button is pressed.
Set Average
Background
An arbitrary number of tests can be averaged to
produce an averaged test result. 1-255 tests can be
averaged. The larger the number of tests that are
averaged, the longer the test time.
Panel operation
1. From the main menu, press MENU, followed
by SETTING and AVGE.
2. Enter the number of number of averages (tests)
using the numerical keys, and then press
ENTER.