Vue de la page 30
BIN FUNCTIONS
61
Set Bin Menu
Set Bin Menu Overview
Bin number Displays the current bin.
Bin Settings Configures the nominal value, nominal units and the
total amount of bins.
SORT BY Chooses the primary or secondary
measurement to sort test results.
Tot_Bin Configures the amount of sort bins.
R_Nom.Val
Q_Nom.Val
Sets the nominal value, depending on the
SORT BY R/Q settings.
Bin parameters Configures the maximum and minimum sort limits for
the current bin.
Max: +
Sets the maximum bin as an absolute
value.
Max: -
Sets the minimum bin as an absolute
value
+: %
Sets the maximum bin value as a positive
percentage offset from the nominal
value.
- %
Sets the minimum bin value as a
negative percentage offset from the
nominal value.
LCR-800 User Manual
62
Q_Max Sets the Maximum Q value for the
current bin.
Q_Min Sets the minimum Q value for the
current bin.
Menu Keys
Scroll up through the menu items
Scroll down through the menu items
Clears the current bin settings.
Goes to the next bin.
Exit the menu
Bin Menu
Background
Before Bin Sorting, the measurement settings must
be configured
Panel operation
1. To access the handler menu, press MENU,
SORT, HANDLER.
2. The Handler menu appears.
SLOW
V : 1.000
RANGE: 1
DELAY : 00000mS
V
EXIT
SET
INT.B
C.V
AVG :
:OFF
1
OFF
MODE
SPEED
CIRCUIT
DISPLAY
BIN SUM
: :VAL
:SER.:R/Q
:FAST
R
24.83
F : 1.0000
Q
MANU
.0000
kHz
SET BIN