PROGRAMMING
107
<value> Test result
Any 7 digit ASCII including sp (+) or
characters and a decimal point.
Primary measurement
value
Example
MAIN:PRIM 32.705<^END>
The primary measurement is 32.705 (primary
measurement unit).
PRIMARY OV01
Measurement
Primary OV01 indicates that the primary measurement exceeds the
measurement range of the LCR meter. For example: If the
impedance of the DUT is less than the measurement range.
Return Syntax PRIM:OV01<^END>
Example PRIM:OV01<^END>
Note, no units are returned
PRIMARY OVER SECONDARY OVER
Measurement
When both the primary and secondary factors exceed the range
(OVER), OVER will be returned.
Return Syntax PRIM:OVER<END>
Example PRIM:OVER<END>
Note, no units are returned
SECONDARY FACTOR & PRIMARY UNIT
Measurement
Returns the secondary measurement results and the primary unit
(R/Q C/D L/Q only). This measurement is the second measurement
displayed after measurements have been started.
Return Syntax MAIN:SECO <value><unit1><^END>
<value> Test Result
Any 6 digit ASCII character including sp (+) or
– characters and a decimal point.
Secondary measurement
value
LCR-800 User Manual
108
<unit1> Primary unit
nF, pF, uF nanofarads, picofarads,
microfarads
k
sp, spsp (sp= space character)
k,
mH, H
sp millihenry, henry
Example
MAIN:SECO .0045nF<^END>
The secondary measurement is .0045 (D) and nF is
the primary measurement unit.
SECONDARY OVER & PRIMARY UNIT
Measurement
Secondary Over indicates that the secondary measurement exceeds
the measurement range of the LCR meter. The unit returned refers to
the primary measurement. Applicable for (R/Q, C/D, L/Q, Z/)
equivalent circuits.
Return Syntax SECO:OVER<unit1><^END>
<unit1> Primary unit
nF, Pf, uF nanofarads, picofarads,
microfarads
ksp, spsp (sp= space character)
k,
mH, H
sp millihenry, henry
Example
SECO:OVER nF<^END>
The secondary measurement is OVER(exceeds range)
and nF is the primary measurement unit.