PEL-2000 Programming Manual
106
BVALue 20
Sets B Value to 20 ohms.
(Low Range only)
AVALue MIN
Sets A Value to the
minimum level for the
specific channel.
AVALue MAX
Sets A Value to the
maximum level for the
specific channel.
Example
:RES:STAT:LOW:BVAL 10
Sets low range CR static
mode B Value to 10
ohms.
Query Syntax
:RESistance:STATic:LOW:AVALue/BVALue?
Return Parameter
<NR2> [MAX|MIN]
Resistance
1 unit = 1 ohm
Returns the resistance of
the A or B Value.
Query Example
:RES:STAT:LOW:AVAL?
MAX
300
Returns the maximum
resistance allowed for
the channel. (PEL-2020)
:RESistance:STATic:LOW:RISE/FALL
Channel Specific
Command
Description
Sets the low range rising/falling slew rate.
Syntax
:RESistance:STATic:LOW:RISE/FALL <NRf+>[A/uS]
Parameter
<NRf+>[A/uS]
Slew rate
RISE/FALL 0.8A/uS
Sets the rising/falling slew
rate to 0.8A/uS
RISE/FALL .8
Sets the rising/falling slew
rate to 0.8 A/uS
RISE/FALL MIN
Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX
Sets to the fastest
rising/falling slew rate.