Query Example
:STAT:CSUM:ENAB?
4
Only the events from
channel 3 can set the
CSUM bit in the Status
Byte Register.
:STATus:CSUMmary:EVENt?
Channel Specific
Command
Description
Returns the status of the Channel Summary Event
register. The Channel Summary Event register is
cleared upon reading.
Query Syntax
:STATus:CSUMmary:EVENt? <NR1>
Return Parameter
<NR1>
Event
<NR1>
Event
1
CH1
16
CH5
2
CH2
32
CH6
4
CH3
64
CH7
8
CH4
128
CH8
Query Example
:STAT:CSUM:EVEN?
4
An event from channel 3
occurred since the last
time the Channel
Summary Event register
was read.
:STATus:QUEStionable:CONDition?
Channel Specific
Command
Description
Returns the status of the Questionable Status
Condition register for the specific channel. See
page 166 for details.
Query Syntax
:STATus:QUEStionable:CONDition? <NR1>
Return Parameter
<NR1>
Condition
<NR1>
Condition
1
OC
16
OT
2
OV
32
G/N
4
OP
64
UVP
8
RV
128~65535
Not used