Query Syntax
:RESistance:STATic:HIGH:AVALue/BVALue?
[MAX|MIN]
Return Parameter
<NR2> [MAX|MIN]
Resistance
1 unit= 1 ohm
Returns the resistance of
the A or B Value.
Query Example
:RES:STAT:HIGH:BVAL?
MAX
15000.0
Returns the maximum
resistance allowed for
the channel for B Value.
(PEL-2020)
:RESistance:STATic:HIGH:RISE/FALL
Channel Specific
Command
Description
Sets the high range rising/falling slew rate.
Syntax
:RESistance:STATic:HIGH:RISE/FALL <NRf+>[A/uS]
Parameter
<NRf+>[A/uS]
Slew rate
RISE/FALL 0.8A/uS
Sets the rising/falling slew
rate to 0.8A/uS
RISE/FALL 0..5
Sets the rising/falling slew
rate to 0.5A/uS
RISE/FALL MIN
Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX
Sets to the fastest
rising/falling slew rate.
Example
:RES:STAT:HIGH:RISE 1.1
Sets the rising slew rate
to 1.1A/uS.
Query Syntax
:RESistance:STATic:HIGH:RISE/FALL? [MIN|MAX]
Return Parameter
<NR2> [MAX|MIN]
Slew rate
1 Unit=1 amp/uS
Returns the slew rate.
MAX/MIN
Returns the allowable
maximum and minimum.