Gw-instek PEL-2000 Series Programming User Manual Manuel d'utilisateur Page 81

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 170
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 80
COMMAND DETAILS
81
:MEASure:POWer?
Channel Specific
Command
Description
This query returns the measured power of the
specific channel.
Query Syntax
:MEASure:POWer? <NR2>
Return Parameter
<NR2> 1 unit = 1 watt
Power at the load input
1.0000
1.0000 watts
Query Example
:MEAS:POW?
1.5
1.5 watts is measured at
the specific channel load
input.
:MEASure:INPut
Channel Specific
Command
Description
This command is for compatibility with other
instruments only and has no action.
Syntax
:MEASure:INPut {LOAD|0| UUT|1}
Parameter
LOAD/0
Disabled
UUT/1
Enabled
Example
:MEAS:INP 0
Disable voltage sense.
Query Syntax
:MEASure:INPut? <NR1>
Return Parameter
<NR1>
Voltage Sense
0
Disabled
1
Enabled
Query Example
:MEAS:INP?
1
Returns the voltage
input status. Voltage
sense is enabled.
Vue de la page 80
1 2 ... 76 77 78 79 80 81 82 83 84 85 86 ... 169 170

Commentaires sur ces manuels

Pas de commentaire