Vue de la page 118
COMMAND DETAILS
119
:SPECification[:PASS]?
Channel Specific
Command
Description
Displays if the Go/NoGo (specification) limit has
passed/failed for the current channel used.
Query Syntax
:SPECification[:PASS]?
Return Parameter
<NR1>
Go/NoGo Specification
0
Fail
1
Pass
Query Example
:SPEC:PASS?
0
Go/NoGo has failed
Query Example
:SPEC?
0
Go/NoGo has failed
:SPECification[:PASS]:CHANnel/
ALLChannel/VOLTage/CURRent?
Channel Specific
Command
Description
Queries if the voltage, current, current channel or
all channels have passed/failed the Go/NoGo
(specification) limits.
VOLTageCC, CR mode,
CURRentCV, CP mode
Query Syntax
:SPECification[:PASS]:CHANnel/ALLChannel/VOLTage
/CURRent?
Return Parameter
<NR1>
Go/NoGo Specification
0
Fail
1
Pass
Query Example
:SPEC:PASS:VOLT?
0
The test has exceeded
the Go/NoGo voltage
limits.
Query Example
:SPEC:VOLT?
0
The test has exceeded
the Go/NoGo voltage
limits.