COMMAND DETAILS
77
Description
Returns the protection levels for electronic load
Query Syntax
:LOAD:PROTection?
Query Example
:LOAD:PROT?
Clears the Channel
Status Register.
Return Parameter
<NR1>
Condition
<NR1>
Condition
1
OC
16
OT
2
OV
32
G/N
4
OP
64
UVP
8
RV
16-128
Not Used
Query Example
:LOAD:PROT?
0
Returns the status of the
Channel Status Register.
Here 0 is returned
indicating no protection
settings have been
tripped.
:LOAD:PROTection:CLEar
Channel Specific
Command
Description
This command clears the Channel Status Register
for the specific channel. See page162.
Syntax
:LOAD:PROTection:CLEar
Example
:LOAD:PROT:CLE
Clears the Channel
Status Register.
:LOAD:TIME?
Channel Specific
Command
Description
This command displays the total load on time. If
the load is on, the load time when the command
was issued is displayed.
Query Syntax
:LOAD:TIME?
Return Parameter
<NR1>1unit = 1 second
Load on time
2.2
2.2 seconds
Vue de la page 76